📞  +353 1 6643768
 
📧  info@medaval.ie
Rossmax Mandaus II (GA101, GA102, GA112)
Device Name:
Mandaus II
Device Model:
GA101 (no stethoscope), GA102 (stethoscope), GA112 (double-head stethoscope)
Manufacturer:
Rossmax International Ltd., 12F - 189 Kang Chien Road, Taipei 114, TAIWAN ROC.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Device Manual:
Description:
The Rossmax Mandaus II (GA101, GA102, GA112) is a hand-held manually-operated digital sphygmomanometer. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for professional use.
Assessment:
The technology used in the Rossmax Mandaus II (GA101, GA102, GA112), to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Legacy approval Older clinical validation; older protocol
BP ESH (Europe) Professional use Published evidence
BP Stride BP Office and hospital use for adults (Unavailable) Published evidence
Validation Publications:

Tasker F, de Greeff A, Liu B, Shennan AH. Validation of a non-mercury digital auscultatory device according to the European Hypertension Society protocol: Rossmax Mandaus II. Blood Press Monit. 2009 Jun;14(3):121-4. doi: 10.1097/MBP.0b013e32832a9dbe. PMID: 19417636.

ESH-IP:2002 - Pass General population