Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations |
BP | ESH (Europe) | Self-measurement | Published evidence |
BP | Stride BP | Self-measurement for adults | Published evidence |
O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. Blood Press Monit. 1996 Feb;1(1):55-61. PMID: 10226203.
BHS:1993 - Pass (B/A) General population
SP10:1992 - Pass General population
Gupta M, Shennan AH, Halligan A, Taylor DJ, de Swiet M. Accuracy of oscillometric blood pressure monitoring in pregnancy and pre-eclampsia. Br J Obstet Gynaecol. 1997 Mar;104(3):350-5. PMID: 9091015.
SP10:1992 - Fail General population
BHS:1993 - Fail (C/D) General population
Artigao LM, Llavador JJ, Puras A, López Abril J, Rubio MM, Torres C, Vidal A, Sanchis C, Divisón JA, Naharro F, Caldevilla D, Fuentes G. Evaluación y validación de los monitores Omron Hem 705 CP y Hem 706/711 para automedidas de presión arterial [Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure.] Aten Primaria. 2000 Feb 15;25(2):74-84. PMID: 10736939. Available from: www.elsevier.es. In Spanish.
BHS:1993 - Pass (A/A) General population
SP10:1987 - Pass General population
Iglesias Bonilla P, Mayoral Sánchez E, Lapetra Peralta J, Iborra Oquendo M, Villalba Alcalá F, Cayuela Domínguez A. Validación de dos sistemas de automedida de presión arterial, modelos OMRON HEM-705 CP y OMRON M1 (HEM 422C2-E) [Validation of two systems of self-measurement of blood pressure, the OMRON HEM-705 CP and OMRON M1 (HEM 422C2-E) models.] Aten Primaria. 2002 Jun 15;30(1):22-8. PMID: 12106576. In Spanish.
BHS:1993 - Pass (A/A) General population
SP10:1987 - Pass General population
Czarkowski M, Mikulska M, Zebrowski M, Baran A, Rózanowski K. Tilt test alters accuracy of the oscillometric device Omron HEM-705CP. J Hum Hypertens. 2004 Apr;18(4):293-4. PMID: 15037881.
Ad Hoc protocol Tilt test
Czarkowski M, Mikulska M, ZEbrowski M, Baran A, Rózanowski K. Venous blood redistribution alters the accuracy of Omron HEM-705CP. Blood Press Monit. 2004 Feb;9(1):9-12. PMID: 15021072.
Ad Hoc protocol Using venous blood redistribution
Furusawa EA, Ruiz MF, Saito MI, Koch VH. Evaluation of the Omron 705-CP blood pressure measuring device for use in adolescents and young adults [Evaluation of the Omron 705-CP blood pressure measuring device for use in adolescents and young adults.] Arq Bras Cardiol. 2005 May;84(5):367-70. Epub: 2005 May 24. PMID: 15917967. In English, Portuguese.
BHS:1993 - Pass (B/A) Adults and adolescents, aged 12 to 21 years (n=60) (Note: Incorrectly stated as A/B in paper)
SP10:2002 - Pass Adults and adolescents, aged 12 to 21 years (n=60)
Vera-Cala LM, Orostegui M, Valencia-Angel LI, López N, Bautista LE. Accuracy of the Omron HEM-705 CP for blood pressure measurement in large epidemiologic studies. Arq Bras Cardiol. 2011 May;96(5):393-8. Epub: 2011 Apr 1. PMID: 21468531. In English, Portuguese, Spanish.
Ad Hoc protocol General population
Aylett M, Marples G, Jones K, Rhodes D. Evaluation of normal and large sphygmomanometer cuffs using the Omron 705CP. J Hum Hypertens. 2001 Feb;15(2):131-4. PMID: 11317193. Available from: www.nature.com.
Medium and large cuffs are compared.
Tansey E, Bell C. Evaluation of digital blood pressure measured with the Omron F3 device as an index of brachial arterial pressure, under different thermal and hormonal conditions. J Hum Hypertens. 2001 Sep;15(9):593-600. PMID: 11550104.
The publication describes the Omron HEM-705CP being used as a reference device for comparison with the Omron F3.
Braam RL, Aslan B, Thien T. Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol. Blood Press Monit. 2004 Feb;9(1):25-30. PMID: 15021075.
Measurements from the Omron HEM-705CP are compared to those from the Omron RX-M, though not in a formal validation.
Cuckson AC, Moran P, Seed P, Reinders A, Shennan AH. Clinical evaluation of an automated oscillometric blood pressure wrist device. Blood Press Monit. 2004 Feb;9(1):31-7. PMID: 15021076.
Measurements from the Omron HEM-705CP are compared to those from the Omron Rx, though not in a formal validation.