Hillrom Home 1700 (H-BP100SBP)
Device Name:
Home 1700
Device Model:
H-BP100SBP
Manufacturer:

OBL: Hill-Rom Holdings Inc., Suite 1000, 130 East Randolph Street, Chicago, IL 60601, UNITED STATES.

OEM: Guangdong Transtek Medical Electronics Co. Ltd., Zone A, 105 Dongli Road, Torch Development District, Zhongshan 528437, CHINA.

Former: Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153, UNITED STATES.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
United States
Device Manual:
Description:
The Hillrom Home 1700 (H-BP100SBP) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Hillrom Home 1700 (H-BP100SBP), to measure blood pressure, has passed in a clinical validation study, in a specific population, according to a recognised standard protocol, as published in a peer-reviewed publication. It has also been declared as being substantially equivalent to that used in another device, according to 510(k), but this is not compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommended in certain condition(s) Validation in specific condition(s). See publications for details.
BP Stride BP Self-measurement for adults (Preferred) Published evidence
BP AMA VDL (US) Office use Published evidence plus 510(k) K161886
BP FDA (US) 510(k) 510(k) K161886
Consumer BodyYear(s)LevelComment
BP Amazon Best Sellers (US) Latest: 2019 A (3.8 / 5)
Device Family:
Validation Publications:

Alpert BS. Validation of the Welch Allyn Home blood pressure monitor with professional SureBP algorithm with a special feature of accuracy during involuntary (tremor) patient movement. Blood Press Monit. 2019 Apr;24(2):89-92. Epub: 2019 Feb 4. doi: 10.1097/MBP.0000000000000365. PMID: 30724792.

81060-2:2013 - Pass Adult subjects whose oscillometric pressures were augmented with simulator-produced tremors. (n=85)

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Hillrom ProBP 2000

Alpert BS. Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing. Blood Press Monit. 2018 Dec;23(6):315-317. Epub: 2018 Oct 11. doi: 10.1097/MBP.0000000000000350. PMID: 30312180.

81060-2:2013 - Pass Adults (n=53) and children (n=35)

Transtek TMB-1776

Liu Z, Chen L. Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults. Blood Press Monit. 2019 Dec;24(6):319-322. Epub: 2019 Sep 26. doi: 10.1097/MBP.0000000000000406. PMID: 31567185.

ESH-IP:2010 - Pass General population

Transtek TMB-986

Liu WJ, Li SG, Song Z, Gong W. Validation of the TRANSTEK blood pressure monitor TMB-986 for home blood pressure monitoring according to the International Protocol. Blood Press Monit. 2010 Oct;15(5):278-80. doi: 10.1097/MBP.0b013e32833e43ca. PMID: 20729726.

ESH-IP:2002 - Pass General population