📞  +353 1 6643768
 
📧  info@medaval.ie
Dinamap ProCare (DPC 100X-EN)
Device Name:
ProCare
Device Model:
DPC 100X-EN
Manufacturer:
GE Medical Systems Ltd. (Dinamap), Amersham Place, Little Chalfont, Buckinghamshire HP7 9NA, UNITED KINGDOM.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for patient monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Single and intermittent measurements
Availability:
Obsolete
Description:
The Dinamap ProCare (DPC 100X-EN) is a patient monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for bedside patient monitoring. This device is now obsolete.
Assessment:
The technology used in the Dinamap ProCare (DPC 100X-EN), to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Legacy approval Older clinical validation; older protocol
BP Stride BP Office and hospital use for adults Published evidence
Device Family:
GE Carescape V100B, Dinamap Procare 420B, Dinamap Carescape V100B, Dinamap ProCare 100B,P, Dinamap ProCare 200B, Dinamap ProCare (DPC 100X-EN)P
Legend: B BIHS Derivative, P Stride Equivalence
Validation Publications:

de Greeff A, Reggiori F, Shennan AH. Clinical assessment of the DINAMAP ProCare monitor in an adult population according to the British Hypertension Society Protocol. Blood Press Monit. 2007 Feb;12(1):51-55. doi: 10.1097/MBP.0b013e3280858b73. PMID: 17303988.

BHS:1993 - Pass (A/A) General population

SP10:1992 - Pass General population

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Dinamap ProCare 100

Reinders A, Reggiori F, Shennan AH. Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population. Blood Press Monit. 2006 Oct;11(5):293-6. PMID: 16932038.

ESH-IP:2002 - Pass General population

Dinamap ProCare 200

Lee CG, Park HM, Shin HJ, Moon JS, Hong YM, Kim NS, Ha IS, Chang MJ, Oh KW. Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents. Korean J Pediatr. 2011 Nov;54(11):463-9. Epub: 2011 Nov 30. doi: 10.3345/kjp.2011.54.11.463. PMID: 22253643. Available from: PMC3254892.

ESH-IP:2010 - Fail Children and adolescents (n=45)