Omron HEM-9200T
Device Model:
HEM-9200T
Manufacturer:
Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Availability according to Countries or Regions:
Japan and United States
Description:
The Omron HEM-9200T is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The technology used in the Omron HEM-9200T, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★ Recommendation Recent clinical validation; recent protocol
BP BIHS (UK and IRL) Self-measurement Published evidence
BP Japanese Society of Hypertension Professional use (2019-2025) Published evidence and a previous equivalence claim to the Omron HEM-7130, HEM-7320F or HEM-7500F
BP Stride BP Self-measurement for adults (Unavailable) (Preferred) Published evidence
BP AMA VDL (US) Self-measurement Published evidence
BP Proprietary declaration of equivalence Manufacturer claim Equivalence to the Omron HEM-7130 (JP) & Omron M6 AC (HEM-7322-E) claim without published evidence
Device Family:
Validation Publications:

Takahashi H, Saito K, Yakura N. Validation of Omron HEM-9200T, a home blood pressure monitoring device for the upper arm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 protocol. J Hum Hypertens. 2022 Apr;36(4):416-419. Epub: 2021 Apr 9. doi: 10.1038/s41371-021-00536-1. PMID: 33837291.

81060-2:2013 - Pass General population

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Omron M2 Intelli IT+ (HEM-7146T2-EBK)

Avagyan A, Zelveian P, Topouchian J, Hakobyan Z, Gharibyan H, Asmar R. Accuracy of the OMRON M2 Intelli IT+ (HEM-7146T2-EBK) device for self-blood pressure measurements in children and adult populations according to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 2:2024) protocol. Blood Press Monit. Epub: 2026 Apr 1. doi: 10.1097/MBP.0000000000000801. PMID: 41919515.

81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population