🏠
Omron M3 (HEM-7200-E)
Device Name:
M3
Device Model:
HEM-7200-E
Manufacturer:
Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Availability according to Countries or Regions:
Ireland
Device Manual:
Description:
The Omron M3 (HEM-7200-E) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The technology used in the Omron M3 (HEM-7200-E), to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommendation Older clinical validation; recent protocol
BP BIHS (UK and IRL) Self-measurement (Previous) Published evidence
BP ESH (Europe) Self-measurement Published evidence
BP Stride BP Self-measurement for adults (Unavailable) Published evidence
Device Family:
Validation Publications:

Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: PMC3237100.

ESH-IP:2010 - Pass General population

Relevant Publications:

Mazoteras Pardo V, Losa Iglesias ME, López Chicharro J, Becerro de Bengoa Vallejo R. The QardioArm App in the Assessment of Blood Pressure and Heart Rate: Reliability and Validity Study. JMIR Mhealth Uhealth. 2017 Dec;5(12):e198. Epub: 2017 Dec 15. doi: 10.2196/mhealth.8458. PMID: 29246880. Available from: PMC5747597.

The use of this device for the reference measurements, in contravention to the ESH-IP 2010 protocol requirements, means that this study can be considered only as a comparison with the Qardio QardioArm A100. It cannot be considered as a validation study.

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Omron M2 (HEM-7117-E)

Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: PMC3237100.

ESH-IP:2002 - Pass General population

Omron M3 Intellisense (HEM-7051-E)

Asmar R, Khabouth J, Topouchian J, El Feghali R, Mattar J. Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Press Monit. 2010 Feb;15(1):49-54. doi: 10.1097/MBP.0b013e3283354b11. PMID: 20032779.

ESH-IP:2002 - Pass General population

Akpolat T, Erdem E, Aydogdu T. Validation of the Omron M3 Intellisense (HEM-7051-E) upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010 in a stage 3-5 chronic kidney disease population. Kidney Blood Press Res. 2012;35(2):82-8. Epub: 2011 Sep 10. doi: 10.1159/000330719. PMID: 21912183.

ESH-IP:2010 - Pass Stage 3-5 chronic kidney disease (Note: Modified protocol)