Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | Legacy approval | Older clinical validation; older protocol |
BP | BIHS (UK and IRL) | Previous recommendation, now archived (Archived) | Published evidence |
BP | ESH (Europe) | Self-measurement | Published evidence |
BP | Stride BP | Self-measurement for adults (Unavailable) | Published evidence |
Legend: B BIHS Derivative, D DET Equivalence, J JHS Implied Equivalence
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2007;3(4):389-400. PMID: 17969368. Available from: PMC2291343.
ESH-IP:2002 - Pass General population
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron HEM Solar (HEM-4500-Sole)
Parati G, Kilama MO, Faini A, Facelli E, Ochen K, Opira C, Mendis S, Wang J, Atkins N, O'Brien E. A new solar-powered blood pressure measuring device for low-resource settings. Hypertension. 2010 Dec;56(6):1047-53. Epub: 2010 Nov 8. doi: 10.1161/HYPERTENSIONAHA.110.160408. PMID: 21059998. Available from: hyper.ahajournals.org.
ESH-IP:2002 - Fail General population (Note: Passed SBP)
Study includes field test in LR settings