Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★★ Recommendation | Recent clinical validation; recent protocol |
BP | BIHS (UK and IRL) | ABPM | Published evidence |
BP | Stride BP | ABPM for adults and children (Preferred) | Published evidence |
BP | MDR Criteria | ABPM | Published evidence |
Hermányi Z, Pokoly B, Visolyi G, Barna I. Evaluation of Meditech ABPM-06 ambulatory blood pressure measuring device, according to the European Society of Hypertension, the British Hypertension Society and the International Organization for Standardization Protocol. Blood Press Monit. 2019 Aug;24(4):208-211. Epub: 2019 May 30. doi: 10.1097/MBP.0000000000000385. PMID: 31162148.
81060-2:2013 - Pass General population
ESH-IP:2010 - Pass General population (Note: See: Blood Press Monit. 2020 Jan;25(1):59-60.)
BHS:1993 - Pass (A/A) General population
Sheppard JP, Lacy P, Chapman N, Clark C, McManus RJ; British and Irish Hypertension Society Blood Pressure Measurement Working Party. Reporting of the Meditech ABPM-06 ambulatory blood pressure device validation study. Blood Press Monit. 2020 Feb;25(1):59-60. doi: 10.1097/MBP.0000000000000425. PMID: 31876570. Available from: journals.lww.com.
This is a letter describing protocol violations in the ESH-IP 2010 validation of the Meditech ABPM-06. See: Blood Press Monit. 2019 Aug;24(4):208-211.