Device Name:
KD-5851
Device Model:
KP-5851
Manufacturer:
Andon Health Co. Ltd., 3 Jinping Road, Ya'an Street, Nankai District, Tianjin 300190, CHINA.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Device Specifications:
Description:
The Andon KD-5851 is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a certified 4-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The accuracy of the technology used in the Andon KD-5851, to measure blood pressure, is certified by Medaval Ltd. It has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★★★ Certified Accuracy Recent clinical validation; recent protocol with certified adherence
BP BIHS (UK and IRL) Self-measurement (Previous) Published evidence
BP ESH (Europe) Self-measurement Published evidence
BP Stride BP Self-measurement for adults (Unavailable) (Preferred) Published evidence
Validation Publications:

Wu L, Jiao Y, Wang C, Chen L, Di D, Zhang H. Validation of the Andon KD-5851 upper arm blood pressure monitor, for self-measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2015 Aug;20(4):228-31. doi: 10.1097/MBP.0000000000000114. PMID: 25768062.

ESH-IP:2010 - Pass General population

Medaval Certification Assessment

Certification ImageMedaval Ltd. Accreditation assessment of the blood pressure measurement technology used in the Andon KD-5851 (KP-5851) upper arm monitor, as validated according to the European Society of Hypertension International Protocol revision 2010. Medical Device Assessment. 2016 Aug 5;2016(1607) 5 p. Epub: 2019 Jan 31. Available from: www.medaval.ie.

With proven validity of validation results according to an accepted protocol, the clinical accuracy of the blood pressure measurement technology, used in the Andon KD-5851, is certified by Medaval Ltd.