Home /  Ypsomed MyLife Pura

Ypsomed MyLife Pura

Device Name:MyLife Pura
Manufacturer:

OBL: Ypsomed Distribution AG, Brunnmattstrasse 6, CH-3401 Burgdorf, SWITZERLAND.

OEM: Bionime Corporation, No 100 - Sec. 2, Daqing Street, South District, Taichung City 40242, TAIWAN PRC.

Measuring functions:Blood glucose
Primary Client Use:Intended for both professional use and self-measurement
Measurement Site:Finger and general Alternative Site Testing
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

The technology used in the Ypsomed MyLife Pura, to measure blood glucose, is among the best rated for accuracy by Medaval Ltd. It has passed in clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★★ RecommendationRecent clinical validation; recent protocol; multiple clinical validations
MDR CriteriaSelf-measurement and professional usePublished evidence

Validation Publications

Baumstark A, Pleus S, Schmid C, Link M, Haug C, Freckmann G. Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1076-86. doi: 10.1177/193229681200600511. PMID: 23063033. Available from: PMC3570841.

15197:2003 General population (Note: 4 test-strip lots; 3 within requirements)

Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.

15197:2003 - Pass General population

15197:2013 - Pass General population

Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: PMC3876330.

15197:2003 - Fail Using insufficient blood samples (n=31)

Freckmann G, Baumstark A, Schmid C, Pleus S, Link M, Haug C. Evaluation of 12 blood glucose monitoring systems for self-testing: system accuracy and measurement reproducibility. Diabetes Technol Ther. 2014 Feb;16(2):113-22. Epub: 2013 Nov 8. doi: 10.1089/dia.2013.0208. PMID: 24205978.

15197:2003 - Pass General population

15197:2013 - Pass General population

Hasslacher C, Kulozik F, Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting. J Diabetes Sci Technol. 2014 May;8(3):466-72. Epub: 2014 Feb 18. doi: 10.1177/1932296814522804. PMID: 24876607. Available from: PMC4455429.

15197:2013 - Pass General population

SKUP. Report from the evaluation SKUP/2010/81: MyLife Pura (Ypsomed, Bionime Corporation), a system for blood glucose measurement. Bergen, Norway: SKUP; 2010. 49 p. Available from: skup.org.

15197:2003 - Pass General population

Relevant Publications

Demircik F, Ramljak S, Hermanns I, Pfützner A, Pfützner A. Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices J Diabetes Sci Technol. 2015 Mar;9(2):262-267. Epub: 2014 Dec 30. doi: 10.1177/1932296814565790. PMID: 25549636. Available from: PMC4604595.

Good accuracy but potential haematocrit influence.