Proton PHCP-868BF
While the technology used in the Proton PHCP-868BF, to measure blood pressure, has been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Accuracy Assessment | Recommendation | Basis | ||
---|---|---|---|---|
Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. | ||
BIHS (UK and IRL) | Self-measurement | Manufacturer declaration of equivalence | ||
ESH (Europe) | Self-measurement | The BIHS recommendation, despite the absence of the scientific evidence required by the same authors' criteria. |
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
H&L HL868BA
Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit. 2008 Oct;13(5):305-8. doi: 10.1097/MBP.0b013e32830d4bd0. PMID: 18799958.
ESH-IP:2002 - Pass General population