Omron MIT Elite (HEM-7300-WE7)
The technology used in the Omron MIT Elite (HEM-7300-WE7), to measure blood pressure, has passed in two clinical validation studies, in pregnancy populations, according to recognised standard protocols, as published in peer-reviewed publications.
Accuracy Assessment | Recommendation | Basis | ||
---|---|---|---|---|
Medaval | Recommended for pregnancy only | Validation in pregnancy | ||
BIHS | Self-measurement and, except in those with large arm circumferrences, during pregnancy | Published evidence, though the paper cited for self measurement refers to a different device |
Chung Y, Brochut MC, de Greeff A, Shennan AH. Clinical accuracy of inflationary oscillometry in pregnancy and pre-eclampsia: Omron-MIT Elite Preg Hyper. 2012 Oct;2(4):411-5. Epub: 2012 May 22. doi: 10.1016/j.preghy.2012.04.001. (Details from: www.pregnancyhypertension.org/article/S2210-7789(12)00030-X/abstract).
BHS:1993 - Pass (A/A) Pregnancy (n=30)
BHS:1993 - Pass (A/A) Pre-eclampsia (n=15)
Exact MIT Elite not specified
James L, Nzelu D, Hay A, Shennan A, Kametas NA. Validation of the Omron MIT Elite blood pressure device in a pregnant population with large arm circumference. Blood Press Monit. 2017 Apr;22(2):109-11. Epub: 2017 Feb 1. doi: 10.1097/MBP.0000000000000239. PMID: 28151753.
BHS:1993 - Fail (D/D) Pregnancy with Obesity
Exact MIT Elite not specified