Omron HEM-7130
The accuracy of the technology used in the Omron HEM-7130, to measure blood pressure, is certified by Medaval Ltd.
Accuracy Assessment | Recommendation | Basis | ||
---|---|---|---|---|
Medaval | ★★★★ Certified Accuracy | Recent clinical validation; recent protocol with certified adherence | ||
BIHS (UK and IRL) | Used as seed device for derivative(s) | |||
ESH (Europe) | Self-measurement | Published evidence | ||
Japanese Society of Hypertension | Self-measurement (2017-2019) | Published evidence | ||
Stride BP | Self-measurement for adults (Preferred) | Published evidence | ||
Consumer Body | Version | Quartile | Comment | |
Choice (Australia) | 2019 | A |
Takahashi H, Yoshika M, Yokoi T. Validation of three automatic devices for the self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: the Omron HEM-7130, HEM-7320F, and HEM-7500F. Blood Press Monit. 2015 Apr;20(2):92-7. doi: 10.1097/MBP.0000000000000096. PMID: 25462531.
ESH-IP:2010 - Pass General population
Medaval Certification Assessment
Medaval Ltd. Accreditation assessment of the blood pressure measurement technology used in the Omron HEM-7130 upper arm monitor, as validated according to the European Society of Hypertension International Protocol revision 2010. Medical Device Assessment. 2016 Aug 5;2016(1618) 4 p. Epub: 2019 Jan 31. Available from:
With proven validity of validation results according to an accepted protocol, the clinical accuracy of the blood pressure measurement technology, used in the Omron HEM-7130, is certified by Medaval Ltd.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron HEM-9700T
Kuwabara M, Harada K, Hishiki Y, Kario K. Validation of an automatic device for the self-measurement of blood pressure in sitting and supine positions according to the ANSI/AAMI/ISO81060-2: 2013 guidelines: the Omron HEM-9700T. Blood Press Monit. Epub: 2019 Feb 14. doi: 10.1097/MBP.0000000000000368.
81060-2:2013 - Pass Sitting position
81060-2:2013 - Pass Supine position