Home /  Omron HEM-7120

Omron HEM-7120

Device Model:HEM-7120
Manufacturer:Sole: Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

While the technology used in the Omron HEM-7120, to measure blood pressure, has been declared as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but the results of the assessment on the other device have not been published either and cannot be verified.

Accuracy AssessmentRecommendationBasis
MedavalNot recommendedThis device has not been clinically validated. Although equivalence to another device is claimed by the manufacturer, this has not been tested scientifically.
Japanese Society of HypertensionSelf-measurement (2017-2018)Implied equivalence to the Omron HEM-7130, HEM-7320F or HEM-7500F