Nissei DM-3000

Device Model:DM-3000
Manufacturer:Nissei Japan Precision Instruments Inc., 2508-13 Nakago Shibukawa, Gunma 377-0293, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for professional use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

The technology used in the Nissei DM-3000, to measure blood pressure, is among the best rated for accuracy by Medaval Ltd. It has passed in two clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★★ RecommendationRecent clinical validation; recent protocol; multiple clinical validations
BIHS (UK and IRL)Professional usePublished evidence
ESH (Europe)Manual and automatic professional usePublished evidence

Validation Publications

Tasker F, De Greeff A, Shennan AH. Development and validation of a blinded hybrid device according to the European Hypertension Society protocol: Nissei DM-3000. J Hum Hypertens. 2010 Sep;24(9):609-16. Epub: 2010 Jan 28. doi: 10.1038/jhh.2009.113. PMID: 20107491. Available from: www.nature.com.

ESH-IP:2002 - Pass General population

Stergiou GS, Karpettas N, Kollias A, Destounis A, Tzamouranis D. A perfect replacement for the mercury sphygmomanometer: the case of the hybrid blood pressure monitor. J Hum Hypertens. 2012 Apr;26(4):220-7. Epub: 2011 Sep 8. doi: 10.1038/jhh.2011.77. PMID: 21900952.

ESH-IP:2010 - Pass General population (Note: Modified protocol)