Home /  Lloyds LBPK1

Lloyds LBPK1

Device Name:LBPK1
Manufacturer:

Brand: Lloyds Pharmacy, Sapphire Court, Walsgrave Triangle, Coventry CV2 2TX, UNITED KINGDOM.

Distributer: Harvard Medical Devices Limited, Unit 1301-02, 13th Floor, Railway Plaza, 39 Chatham Road South, Tsimshatsui, Kowloon, HONG KONG PRC.

Other: Kinetik Medical Devices Limited, Lockeridge House, Lockeridge, Marlborough, Wiltshire SN8 4EL, UNITED KINGDOM.

Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

While the technology used in the Lloyds LBPK1, to measure blood pressure, has been declared as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but the results of the assessment on which the declaration has been based have not been published either and cannot be verified. However, there are validation studies proving clinical validation of the same measurement technology, according to other equivalence claims.

Accuracy AssessmentRecommendationBasis
MedavalNoneAlthough equivalence to another device is claimed, this has not been tested to MDR requirements.
BIHS (UK and IRL)Previous recommendation, now archivedManufacturer declaration of equivalence to a device without published evidence
ESH (Europe)Self-measurementThe BIHS recommendation, despite the absence of the scientific evidence required by the same authors' criteria.
 
Consumer BodyVersionQuartileComment
Which? (UK)PreviousA 

Device Family:Kinetik BPM1CB, Kinetik BPM1KB, Kinetik BPM1KTLB, Lloyds LBP1B, Lloyds LBPK1B (B BIHS Derivative)