IEM Mobil-O-Graph NG

Device Name:Mobil-O-Graph NG
Manufacturer:I.E.M. GmbH (Industrielle Entwicklung Medizintechnik und Vertriebsgesellschaft GmbH), Cockerillstraße 69, D-52222 Stolberg, GERMANY.
Measuring functions:Blood pressure
Primary Client Use:Intended for ambulatory blood pressure monitoring
Measurement Site:Upper Arm
Measurement Occurrence:Intermittent measurements at specified intervals or times
Availability:Available Currently
Assessment:

The technology used in the IEM Mobil-O-Graph NG, to measure blood pressure, has passed in three clinical validation studies, in general populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★ RecommendationRecent clinical validation; older protocol; multiple clinical validations
BIHS (UK and IRL)Published evidence
ESH (Europe)ABPMPublished evidence

Device Family:Cardioline Walk200bB, Medset Scanlight-IIIB, Welch Allyn ABPM 7100B (B BIHS Derivative)
Validation Publications

Jones CR, Taylor K, Chowienczyk P, Poston L, Shennan AH. A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor. Blood Press Monit. 2000 Aug;5(4):233-8. PMID: 11035866.

BHS:1993 - Pass (B/A) General population

SP10:1992 - Pass General population

Franssen PM, Imholz BP. Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria. Blood Press Monit. 2010 Aug;15(4):229-31. PMID: 20658764.

ESH-IP:2002 - Pass General population

Weiß W, Tölle M, Zidek W, van der Giet M. Validation of the mobil-O-Graph: 24 h-blood pressure measurement device. Blood Press Monit. 2010 Aug;15(4):225-8. doi: 10.1097/MBP.0b013e328338892f. PMID: 20216407.

BHS:1993 - Pass General population

Relevant Publications

Weiss W, Gohlisch C, Harsch-Gladisch C, Tölle M, Zidek W, van der Giet M. Oscillometric estimation of central blood pressure: validation of the Mobil-O-Graph in comparison with the SphygmoCor device. Blood Press Monit. 2012 Jun;17(3):128-31. doi: 10.1097/MBP.0b013e328353ff63. PMID: 22561735.

Measurements from the IEM Mobil-O-Graph NG are compared to those from the Sphygmocor, though not in a formal validation.

Kallem RR, Meyers KE, Sawinski DL, Townsend RR. A comparison of two ambulatory blood pressure monitors worn at the same time. J Clin Hypertens (Greenwich). 2013 May;15(5):321-5. Epub: 2013 Apr 8. doi: 10.1111/jch.12098. PMID: 23614845. Available from: onlinelibrary.wiley.com.

Measurements from the IEM Mobil-O-Graph NG are compared to those from the SL-90207, though not in a formal validation.

Kallem RR, Meyers KE, Cucchiara AJ, Sawinski DL, Townsend RR. Blood pressure variability of two ambulatory blood pressure monitors. Blood Press Monit. 2014 Apr;19(2):98-102. doi: 10.1097/MBP.0000000000000019. PMID: 24374722.

Measurements from the IEM Mobil-O-Graph NG are compared to those from the SL-90207, though not in a formal validation.

Sarafidis PA, Lazaridis AA, Imprialos KP, Georgianos PI, Avranas KA, Protogerou AD, Doumas MN, Athyros VG, Karagiannis AI. A comparison study of brachial blood pressure recorded with Spacelabs 90217A and Mobil-O-Graph NG devices under static and ambulatory conditions. J Hum Hypertens. 2016 Dec;30(12):742-9. Epub: 2016 Mar 3. doi: 10.1038/jhh.2016.11. PMID: 26935286.

Measurements from the IEM Mobil-O-Graph NG are compared to those from the SL-90207A and a standard mercury sphygmomanometer, though not in a formal validation.