Home /  Fukuda Denshi DS-7200

Fukuda Denshi DS-7200

Device Name:Dynascope
Device Model:DS-7200
Manufacturer:Sole: Fukuda Denshi Co. Ltd., 2-23-5 Kita Ueno, Taito-Ku, Tokyo 110-0014, JAPAN.
Measuring functions:Blood pressure, Arrhythmias, SpO2, Temperature, ECG, Respiration
Primary Client Use:Intended for patient monitoring
Measurement Site:Upper Arm
Measurement Occurrence:Intermittent measurements at specified intervals or times
Availability:Available Currently
Assessment:

While the technology used in the Fukuda Denshi DS-7200, to measure blood pressure, has been claimed as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but there is no evidence provided on a validation of the other device.

Accuracy AssessmentRecommendationBasis
MedavalNoneAlthough equivalence to another device is claimed, this has not been tested to MDR requirements.
Japanese Society of HypertensionProfessional use (2016, 2018-2019)Implied equivalence to the Fukuda Denshi DS-7000

Validation Publications for Equivalent Devices

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Fukuda Denshi DS-7000

Alpert BS, Blakely DW. Validation of the Fukuda Denshi DS-7000 patient monitor by AAMI standard testing. Blood Press Monit. 2009 Dec;14(6):274-6. doi: 10.1097/MBP.0b013e328332fd6d. PMID: 19923985.

BHS:1993 - Pass (A/A) In Normal Mode

BHS:1993 - Pass (A/A) In Quick Mode

SP10:2002 - Pass Adults and Children - In Normal Mode

SP10:2002 - Pass Adult and Children - In Quick Mode

SP10:2002 - Pass Infants (n=15) and Neonate (n=18)