Assessment:

While the technology used in the Citizen CHU-302, to measure blood pressure, has been claimed as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but there is no evidence provided on a validation of the other device.

Accuracy AssessmentRecommendationBasis
MedavalNoneAlthough equivalence to another device is claimed, this has not been tested to MDR requirements.
Japanese Society of HypertensionSelf-measurement (2016-2018)Unpublished equivalence to Citizen CH-432B, by questionnaire

Device Family:Citizen CH-432J, Citizen CH-432BD, Citizen CH-452D,J, Citizen CH-453FJ, Citizen CH-550J, Citizen CHU-302J, Citizen CHU-501J, Citizen CHU-502J (D DET Equivalence, J JHS Implied Equivalence)
Validation Publications for Equivalent Devices

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Citizen CH-432B

Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C. Blood Press Monit. 2008 Feb;13(1):55-62. doi: 10.1097/MBP.0b013e3282cb599a. PMID: 18199925. Erratum in: Blood Press Monit. 2008 Aug;13(4):243.