A&D UB-510
The technology used in the A&D UB-510, to measure blood pressure, has been compared to a clinically validated device but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Accuracy Assessment | Recommendation | Basis | ||
---|---|---|---|---|
Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. | ||
Japanese Society of Hypertension | Self-measurement (2018) | Implied equivalence to the A&D UB-511 or UB-512 |
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
A&D UB-511
Angeli F, Sardone M, Angeli E, Repaci S, Gattobigio R, Verdecchia P. Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure. Blood Press Monit. 2006 Dec;11(6):349-54. PMID: 17106320.
BHS:1993 - Pass (B/B) General population
A&D UB-512
Angeli F, Sardone M, Angeli E, Repaci S, Gattobigio R, Verdecchia P. Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure. Blood Press Monit. 2006 Dec;11(6):349-54. PMID: 17106320.
BHS:1993 - Pass (B/B) General population