A&D TM-2657VP

Device Model:TM-2657VP
Manufacturer:Sole: A&D Company Ltd., 3-23-14 Higashi-Ikebukuro, Toshima-Ku, 170-0013 Tokyo, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for use as a public facility
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

While the technology used in the A&D TM-2657VP, to measure blood pressure, has been claimed as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but there is no evidence provided on a validation of the other device.

Accuracy AssessmentRecommendationBasis
MedavalNoneAlthough equivalence to another device is claimed, this has not been tested to MDR requirements.
Japanese Society of HypertensionProfessional use (2019)Implied equivalence to the A&D TM-2656

Device Family:A&D TM-2656J, A&D TM-2656VPWJ, A&D TM-2657J, A&D TM-2657PJ, A&D TM-2657VPJ (J JHS Implied Equivalence)
Validation Publications for Equivalent Devices

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

A&D TM-2656

Zeng WF, Liu M, Kang YY, Li Y, Wang JG. Validation of the fully automated A&D TM-2656 blood pressure monitor according to the British Hypertension Society Protocol. Blood Press Monit. 2013 Aug;18(4):223-6. doi: 10.1097/MBP.0b013e3283624abd. PMID: 23681206.

BHS:1993 - Pass (A/A) General population