Hillrom ProBP 2000
Manufacturer:

OBL: Hill-Rom Holdings Inc., Suite 1000, 130 East Randolph Street, Chicago, IL 60601, UNITED STATES.

Former: Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153, UNITED STATES.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
United States
Description:
The Hillrom ProBP 2000 is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood pressure measurements are taken from the upper arm and it is intended for professional use.
Assessment:
The technology used in the Hillrom ProBP 2000, to measure blood pressure, has passed in a clinical validation study, in a general population including children, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
Medaval ★★ Recommendation Recent clinical validation; recent protocol
Stride BP Office use for adults and children (Preferred) Published evidence
AMA VDL (US) Professional use Published evidence
MDR Criteria Professional use Published evidence
Device Family:
Validation Publications:

Alpert BS. Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing. Blood Press Monit. 2018 Dec;23(6):315-317. Epub: 2018 Oct 11. doi: 10.1097/MBP.0000000000000350. PMID: 30312180.

81060-2:2013 - Pass Adults (n=53) and children (n=35)

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Hillrom Home 1700 (H-BP100SBP)

Alpert BS. Validation of the Welch Allyn Home blood pressure monitor with professional SureBP algorithm with a special feature of accuracy during involuntary (tremor) patient movement. Blood Press Monit. 2019 Apr;24(2):89-92. Epub: 2019 Feb 4. doi: 10.1097/MBP.0000000000000365. PMID: 30724792.

81060-2:2013 - Pass Adult subjects whose oscillometric pressures were augmented with simulator-produced tremors. (n=85)

Transtek TMB-986

Liu WJ, Li SG, Song Z, Gong W. Validation of the TRANSTEK blood pressure monitor TMB-986 for home blood pressure monitoring according to the International Protocol. Blood Press Monit. 2010 Oct;15(5):278-80. doi: 10.1097/MBP.0b013e32833e43ca. PMID: 20729726.

ESH-IP:2002 - Pass General population