Device Model:
DS-400
Manufacturer:
Nihon Seimitsu Sokki Co. Ltd. (Nissei), 2508-13 Nakago Shibukawa, Gunma 377-0293, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Device Manual:
Description:
The Nissei DS-400 is a hand-held semi-automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating. Blood pressure measurements are taken from the upper arm and it is intended for self-measurement and home use.
Assessment:
The technology used in the Nissei DS-400, to measure blood pressure, has passed in two clinical validation studies, between general, pregnancy and other specific populations, according to recognised standard protocols, as published in peer-reviewed publications.
Recommendations:
Accuracy AssessmentRecommendationBasis
Medaval Recommendation Older clinical validation; older protocol; multiple clinical validations
ESH (Europe) Self-measurement for adults and pregnancy Published evidence
Stride BP Self-measurement for adults Published evidence
Device Family:
Beurer BM 60D , Nissei DS-400D , Nissei DS-500D
Legend: D DET Equivalence
Validation Publications:

Duhig KE, De Greeff A, Van Der Westhuizen A, Baker E, Shennan AH. Validation of the Nissei DS-400 in a low-resource setting. Blood Press Monit. 2009 Jun;14(3):132-5. doi: 10.1097/MBP.0b013e32832ce136. PMID: 19433979.

ESH-IP:2002 - Pass Low resource setting

de Greeff A, Shennan AH. Clinical accuracy of a low cost portable blood pressure device in pregnancy and pre-eclampsia: the Nissei DS-400. Trop Doct. 2015 Jul;45(3):168-73. Epub: 2015 Apr 23. doi: 10.1177/0049475515581542. PMID: 25911652.

BHS:1993 - Pass (A/A) Pregnancy and pre-eclampsia (n=45)

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Beurer BM 60

Tholl U, Lüders S, Bramlage P, Dechend R, Eckert S, Mengden T, Nürnberger J, Sanner B, Anlauf M. The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control. Blood Press Monit. 2016 Aug;22(4):197-205. doi: 10.1097/MBP.0000000000000186. PMID: 26998590.

DHL:1999 - Pass General population (Note: 2011)