Device Name:
Vital Signs
          Manufacturer:
Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153, UNITED STATES.
          Measuring functions:
Blood Pressure and PR
          Primary Client Use:
Intended for patient monitoring
          Measurement Site:
Upper Arm
          Measurement Occurrence:
Single measurements only
          Availability:
Obsolete
          Device Manual:
         
       
      
        Description:
The Welch Allyn Vital Signs is a vital signs monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for bedside patient monitoring. This device is now obsolete.
        Assessment:
The technology used in the Welch Allyn Vital Signs, to measure blood pressure, has passed in two clinical validation studies, in general populations, according to recognised standard protocols, as published in peer-reviewed publications.
        Recommendations:
        
          
            |  | Accuracy Assessment | Recommendation | Basis | 
            
              | BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations | 
            
              | BP | Stride BP | Office use for adults and pregnancy | Published evidence | 
            
              | BP | FDA (US) | 510(k) | 510(k) K171621 | 
          
         
       
      
        
          Validation Publications:
          
            Jones CR, Taylor K, Poston L, Shennan AH. Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor. J Hum Hypertens. 2001 Mar;15(3):191-5. PMID: 11317204.
            BHS:1993 - Pass (A/A) General population
            SP10:1996 - Pass General population
            Reinders A, Cuckson AC, Jones CR, Poet R, O'Sullivan G, Shennan AH. Validation of the Welch Allyn 'Vital Signs' blood pressure measurement device in pregnancy and pre-eclampsia. BJOG. 2003 Feb;110(2):134-8. PMID: 12618156.
            BHS:1993 - Pass General population
            Wong SN, Tz Sung RY, Leung LC. Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children. Blood Press Monit. 2006 Oct;11(5):281-91. PMID: 16932037.
            ESH-IP:2002 - Fail General population