Transtek TMB-1491-S
Device Model:
TMB-1491-S
Manufacturer:
Guangdong Transtek Medical Electronics Co. Ltd., Zone A, 105 Dongli Road, Torch Development District, Zhongshan 528437, CHINA.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Description:
The Transtek TMB-1491-S is an automatic blood pressure monitor. Medaval has not found evidence proving the accuracy of its blood pressure measurement technology. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
There appears to be no peer-reviewed clinical validation information available on the technology used in the Transtek TMB-1491-S to measure blood pressure.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Not recommended This device has not been clinically validated.
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Transtek TMB-1491

Tian H, Zeng S, Zhong X, Gong W, Liu W. Validation of Transtek blood pressure monitor TMB-1491 for self-measurement according to the European Society of Hypertension International Protocol reversion 2010. Blood Press Monit. 2015 Oct;20(5):280-2. Epub: 2015 May 5. doi: 10.1097/MBP.0000000000000125. PMID: 25968092.

ESH-IP:2010 - Pass General population