📞  +353 1 6643768
 
📧  info@medaval.ie
Terumo ES-P2000
Device Model:
ES-P2000
Manufacturer:
Terumo Corporation, 2-44-1 Hatagaya, Shibuya-ku, Tokyo 151-0072, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for use as a public facility
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Description:
The Terumo ES-P2000 is a waiting-room blood pressure monitor. Although its blood pressure measurement technology has been evaluated, it is not possible to base a determination on its accuracy from this research. Blood pressure measurements are taken from the upper arm. It is intended for use as a public facility.
Assessment:
The technology used in the Terumo ES-P2000, to measure blood pressure, has been assessed, in a general population, according to a non-standard protocol.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol.
BP Japanese Society of Hypertension Self-measurement (Previous: 2016-2019) Incomplete internal report
Device Family:
Terumo ES-P2000J, Terumo ES-P2020DZJ, Terumo ES-P2020ZZJ
Legend: J JHS Implied Equivalence
Validation Publications:

血圧計に関するワーキンググループ, 日本高血圧学会学術委員会 [Working Group on Sphygmomanometers, Academic Committee on Hypertension in Japan] [Test reports on the Terumo R800 and P2000] [Internet] Tokyo: JSH; 2010. Available from: www.jpnsh.jp.

JIS T 1115:2005 General population

JIS T 1115:2005 General population

There is insufficient information provided in either of two reports to enable scientific assessment.