TaiDoc TD-4235B GlucoRx Q
Device Name:
GlucoRx Q
Device Model:
TD-4235B
Manufacturer:

Distributer: GlucoRx Ltd., 4 Wintonlea, Monument Way West, Woking, Surrey GU21 5EN, UNITED KINGDOM.

OEM: TaiDoc Technology Corporation, 6F - 127 Wugong 2nd Road, 24888 Wugu District, New Taipei City, TAIWAN ROC.

Measuring functions:
Blood glucose
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Finger and general Alternative Site Testing
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The TaiDoc TD-4235B GlucoRx Q is an automatic blood glucose meter. Its blood glucose measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood glucose measurements are normally taken from the finger but alternative site testing is allowed. It is intended for self-measurement and home use.
Assessment:
The technology used in the TaiDoc TD-4235B GlucoRx Q, to measure blood glucose, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BG Medaval ★★ Recommendation Recent clinical validation; recent protocol
BG MDR Criteria Self-measurement Published evidence
Validation Publications:

Thakur P. Accuracy Evaluation of a Blood Glucose Meter System for Self-Testing with Three Strip Lots Following ISO 15197:2013 and EN ISO 15197:2015 Rep Dis Mark. 2017 Nov 13;1(1):101 (2 p). Available from: www.omicsonline.org. Letter.

15197:2013 - Pass General population