Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations |
BP | BIHS (UK and IRL) | ABPM | Published evidence |
BP | ESH (Europe) | ABPM | Published evidence |
BP | Stride BP | ABPM for adults | Published evidence |
Baumgart P, Kamp J. Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor. Blood Press Monit. 1998 Oct;3(5):303-307. PMID: 10212370.
BHS:1993 - Pass (A/A) General population
SP10:1992 - Pass General population (Note: Standing, sitting and supine positions)
Redwine KM, James LP, O’Riordan MA, Sullivan JE, Blumer JL; for the Network of Pediatric Pharmacology Research Units. Accuracy of the Spacelabs 90217 ambulatory blood pressure monitor in a pediatric population. Blood Press Monit. 2015 Oct;20(5):295-8. doi: 10.1097/MBP.0000000000000132. PMID: 26529437.
BHS:1993 - Pass (A/B) Children, aged 6 to 17 years (n=112)
Amoore JN, Dewar D, Gough K, Padfield PL. Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? Blood Press Monit. 2005 Feb;10(1):51-6. PMID: 15687874.
Measurements from the Spacelabs 90217 are compared to those from a simulator, though not in a formal validation.
Pang TC, Brown MA. Accuracy of ambulatory blood pressure monitors in routine clinical practice. Am J Hypertens. 2006 Aug;19(8):801-9. PMID: 16876678.
Measurements from the Spacelabs 90217 are compared to those from a standard mercury sphygmomanometer (retrospectively), though not in a formal validation.
Nyvad J, Christensen KL, Buus NH, Reinhard M. The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring. J Clin Hypertens (Greenwich). 2021 Jan;23(1):61-70. Epub: 2020 Dec 17. doi: 10.1111/jch.14135. PMID: 33350030.
Measurements from the Spacelabs 90217 are compared to those from the Somno Somnotouch NIBP (TON100), though not in a formal validation.