Spacelabs 90207
Device Model:
90207
Manufacturer:
Spacelabs Healthcare, 35301 SE Center St., Snoqualmie, WA 98065, UNITED STATES.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for ambulatory blood pressure monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Available Currently
Availability according to Countries or Regions:
Ireland
Device Manual:
Description:
The Spacelabs 90207 is an ambulatory blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for ambulatory blood pressure monitoring.
Assessment:
The technology used in the Spacelabs 90207, to measure blood pressure, has passed in six clinical validation studies, between general, pregnancy and other specific populations, according to recognised standard protocols, as published in peer-reviewed publications.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommendation Older clinical validation; older protocol; multiple clinical validations
BP BIHS (UK and IRL) ABPM Published evidence
BP ESH (Europe) ABPM Published evidence
BP Stride BP ABPM for adults (Unavailable) Published evidence
Validation Publications:

Groppelli A, Omboni S, Ravogli A, Villani A, Parati G, Mancia G. Validation of the SpaceLabs 90202 and 90207 devices for ambulatory blood pressure monitoring by comparison with intra-arterial resting and ambulatory measurements. J Hypertens. 1991 Dec;9(Suppl 6):S334-5. PMID: 1818988.

IA protocol General population (Note: No abstract online)

O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol. J Hypertens. 1991;9(Suppl 5):S25-S31.

BHS:1990 - Pass (B/B) General population

SP10:1987 - Pass General population

Groppelli A, Omboni S, Parati G, Mancia G. Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure. Hypertension. 1992 Aug;20(2):227-32. PMID: 1639465.

Ad Hoc protocol General population (Note: IA validation)

O'Brien E, Mee F, Atkins N, Halligan A, O'Malley K. Accuracy of the SpaceLabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society protocol. J Hypertens. 1993 Dec;11(Suppl 5):S282-3. PMID: 8158386.

BHS:1990 - Fail (A/C) Pregnancy

Shennan AH, Kissane J, de Swiet M. Validation of the SpaceLabs 90207 ambulatory blood pressure monitor for use in pregnancy. Br J Obstet Gynaecol. 1993 Oct;100(10):904-8. PMID: 8217971.

BHS:1993 - Fail (B/C) Pregnancy

Vine S. Validation of the SpaceLabs 90207 ambulatory blood pressure monitor for use in pregnancy. Br J Obstet Gynaecol. 1994 May;101(5):463. PMID: 8018630.

BHS:1993 Pregnancy (Note: No abstract online)

Mayoral Sánchez E, Díez Naz AD, Lapetra Peralta J, Santos Lozano JM, García de la Corte F, Rodríguez-Morcillo Guardia A. Validación del sistema de monitorización ambulatoria de presión arterial modelo SpaceLabs 90207. [An evaluation of the Spacelabs 90207 model of the ambulatory arterial pressure monitoring system.] Med Clin (Barc). 1994 Sep 24;103(9):326-30. PMID: 7967890. In Spanish.

BHS:1993 - Pass (A/B) General population

Brown MA, Buddle ML, Bennett M, Smith B, Morris R, Whitworth JA. Ambulatory blood pressure in pregnancy: comparison of the Spacelabs 90207 and Accutracker II monitors with intraarterial recordings. Am J Obstet Gynecol. 1995 Jul;173(1):218-23. PMID: 7631686.

SP10:1992 - Fail Pregnancy (Note: IA validation (n=25), Failed DBP)

Belsha CW, Wells TG, Bowe Rice H, Neaville WA, Berry PL. Accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit. 1996 Apr;1(2):127-133. PMID: 10226214.

BHS:1990 - Fail (C/D) Adults and children

SP10:1992 - Fail Adults and chlidren

Iqbal P, Fotherby MD, Potter JF. Validation of the SpaceLabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit. 1996 Aug;1(4):367-373. PMID: 10226261.

BHS:1993 - Pass (A/A) Elderly

Shennan A, Halligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measurements in severe pre-eclampsia: validation of the SpaceLabs 90207. Br J Obstet Gynaecol. 1996 Feb;103(2):171-3. Epub: 2005 Aug 19. doi: 10.1111/j.1471-0528.1996.tb09671.x. PMID: 8616136. (Details from: onlinelibrary.wiley.com). Correspondance in: Br J Obstet Gynaecol. 1996 Jul;103(7):721-2.

BHS:1993 Severe pre-eclampsia

No abstract available online

Márquez Contreras E, López de Andrés M, Casado Martínez JJ, Martín de Pablos JL, Moreno García JP, López Zamorano JM. Validación del monitor automático no invasivo de monitorización ambulatoria de la presión arterial Spacelabs 90207. [Evaluation of Spacelabs 90207, the noninvasive automatic monitor for ambulatory monitoring of blood pressure.] Aten Primaria. 1998 Feb 15;21(2):105-8. PMID: 9608135. In Spanish.

SP10:1987 - Pass General population

BHS:1990 - Pass (B/B) General population

Peixoto AJ, Gray TA, Crowley ST. Validation of the SpaceLabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit. 1999 Oct;4(5):217-21. PMID: 10547641.

BHS:1993 - Fail (C/B) During haemodialysis

Czarkowski M, Mikulska M, Baran A, Zebrowski M, Rózanowski K. Accuracy of SpaceLabs 90207 is altered by venous blood redistribution. Blood Press. 2003;12(5-6):334-9. PMID: 14763666.

Ad Hoc protocol Using venous blood redistribution

Elvan-Taşpinar A, Uiterkamp LA, Sikkema JM, Bots ML, Koomans HA, Bruinse HW, Franx A. Validation and use of the Finometer for blood pressure measurement in normal, hypertensive and pre-eclamptic pregnancy. J Hypertens. 2003 Nov;21(11):2053-60. PMID: 14597848.

BHS:1993 - Pass (A/A) Normotensive pregnancy (n=54) (Note: SL-90207 validated in conjunction with Finometer)

BHS:1993 - Fail (B/C) Pre-eclampsia (n=31)

BHS:1993 - Pass (B/A) Hypertensive pregnancy (n=38)

Miszkowska-Nagórna E, Neubauer-Geryk J, Wolf J, Wielicka M, Raczak G, Narkiewicz K, Bieniaszewski L. The accuracy of SpaceLabs 90207 in blood pressure monitoring in patients with atrial fibrillation. Blood Press. 2018 Feb;27(1):3-9. Epub: 2017 Jul 24. doi: 10.1080/08037051.2017.1356200. PMID: 28738693.

Ad Hoc protocol Atrial fibrillation (n=47), Post cardioversion (n=29)

Relevant Publications:

O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207, Novacor DIASYS 200, Del Mar Avionics Pressurometer IV and Takeda TM-2420 ambulatory systems according to British and American criteria. J Hypertens. 1991 Dec;9(Suppl 6):S332-3. PMID: 1818987.

This is a supplementary presentation of validation results for the Spacelabs 90207.

See: O'Brien E et al. 1991

O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol. J Hypertens. 1991 Jun;9(6):573-4. PMID: 1653299. (Details from: journals.lww.com).

This is a supplementary presentation of validation results for the Spacelabs 90207.

See: O'Brien E et al. 1991

O'Brien E, Atkins N, Mee F, O'Malley K. Comparative accuracy of six ambulatory devices according to blood pressure levels. J Hypertens. 1993 Jun;11(6):673-5. PMID: 8397247.

This is a supplementary presentation of validation results for the Spacelabs 90207.

See: O'Brien E et al. 1991

Pirie AM, Quinn M. Oscilometric blood pressure measurements in severe pre-eclampsia: validation of SpaceLabs 90207. Br J Obstet Gynaecol. 1996 Jul;103(7):721-2. Epub: 2005 Aug 19. doi: 10.1111/j.1471-0528.1996.tb09851.x. PMID: 8688408. Available from: onlinelibrary.wiley.com. Correspondance re: Br J Obstet Gynaecol. 1996 Feb;103(2):721-3.

The publication is a letter or correspondence in which the Spacelabs 90207 is referenced.

Amoore JN, Geake WB. Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator. J Hum Hypertens. 1997 Mar;11(3):163-9. PMID: 9175568.

Measurements from the Spacelabs 90207 are compared to those from a simulator, though not in a formal validation.

Amoore JN, Dewar D, Gough K, Padfield PL. Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? Blood Press Monit. 2005 Feb;10(1):51-6. PMID: 15687874.

Measurements from the Spacelabs 90207 are compared to those from a simulator, though not in a formal validation.

Pang TC, Brown MA. Accuracy of ambulatory blood pressure monitors in routine clinical practice. Am J Hypertens. 2006 Aug;19(8):801-9. PMID: 16876678.

Measurements from the Spacelabs 90207 are compared to those from a standard mercury sphygmomanometer (retrospectively), though not in a formal validation.

Kallem RR, Meyers KE, Sawinski DL, Townsend RR. A comparison of two ambulatory blood pressure monitors worn at the same time. J Clin Hypertens (Greenwich). 2013 May;15(5):321-5. Epub: 2013 Apr 8. doi: 10.1111/jch.12098. PMID: 23614845. Available from: onlinelibrary.wiley.com.

Measurements from the Spacelabs 90207 are compared to those from the IEM Mobil-O-Graph, though not in a formal validation.

Kallem RR, Meyers KE, Cucchiara AJ, Sawinski DL, Townsend RR. Blood pressure variability of two ambulatory blood pressure monitors. Blood Press Monit. 2014 Apr;19(2):98-102. doi: 10.1097/MBP.0000000000000019. PMID: 24374722.

Measurements from the Spacelabs 90207 are compared to those from the IEM Mobil-O-Graph, though not in a formal validation.