Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★★ Recommendation | Recent clinical validation; recent protocol |
BP | Stride BP | ABPM for adults (Previous) | Published evidence |
Bilo G, Zorzi C, Ochoa Munera JE, Torlasco C, Giuli V, Parati G. Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2015 Oct;20(5):291-4. Epub: 2015 Apr 28. doi: 10.1097/MBP.0000000000000124. PMID: 25932885.
ESH-IP:2010 - Pass General population
Manning LS, Robinson TG, Panerai RB. The SOMNOtouch device as a novel method for measuring short-term blood pressure variability: a comparison with the Finometer. Blood Press Monit. 2015 Dec;20(6):361-8. doi: 10.1097/MBP.0000000000000128. PMID: 26154853.
Measurements from the Somno Somnotouch NIBP (TON100) are compared to those from the Finapres Finometer Midi, though not in a formal validation.
Golberg M, Ruiz-Rivas J, Polani S, Beiderman Y, Zalevsky Z. Large-scale clinical validation of noncontact and continuous extraction of blood pressure via multipoint defocused photonic imaging. Appl Opt. 2018 Mar 1;57(7):B45-B51. doi: 10.1364/AO.57.000B45. PMID: 29522038.
The publication describes the Somno Somnotouch NIBP (TON100) being used as a reference device for comparison with .
Nyvad J, Christensen KL, Buus NH, Reinhard M. The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring. J Clin Hypertens (Greenwich). 2021 Jan;23(1):61-70. Epub: 2020 Dec 17. doi: 10.1111/jch.14135. PMID: 33350030.
Measurements from the Somno Somnotouch NIBP are compared to those from the Spacelabs SL-90217, though not in a formal validation.
Bothe TL, Hulpke-Wette M, Barbarics B, Patzak A, Pilz N. Accuracy of cuff-less, continuous, and non-invasive blood pressure measurement in 24-h ABPM in children aged 5-17. Blood Press. 2023 Dec;32(1):2255704. doi: 10.1080/08037051.2023.2255704. PMID: 37683118.
Measurements from the Somno Somnotouch NIBP (TON100) are compared to those from the IEM Mobil-O-Graph, though not in a formal validation.