Sejoy Joytech DBP-6279B
Device Model:
DBP-6279B
Manufacturer:
Hangjhou Sejoy Electronics & Instrument Co. Ltd. Joytech Healthcare Co. Ltd., Area C Building 2, 365 Wuzhou Road, Yuhang Economic Development Zone, Yuhang District, Hangzhou City, Zhejiang 311100, CHINA.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Connected Health Technologies:
Bluetooth
Availability:
Not made available
Description:
The Sejoy Joytech DBP-6279B is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 3-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Sejoy Joytech DBP-6279B, to measure blood pressure, is among the best rated for accuracy by Medaval Ltd. It has passed in two clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.
Recommendations:
| Accuracy Assessment | Recommendation | Basis | |
| BP | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
| BP | Stride BP | Self-measurement for adults (Preferred) | Published evidence |
| BP | AMA VDL (US) | Self-measurement for adults | Published evidence |
Device Family:
Aposán DBP-6194P, Motif Medical BAB0001P, Sejoy Joytech DBP-6179P, Sejoy Joytech DBP-6191P, Sejoy Joytech DBP-6192P, Sejoy Joytech DBP-6279BP, Sejoy Joytech DBP-6191HDP, Sejoy Joytech DBP-6282BP, Sejoy Joytech DBP-6285BP
Legend: P Stride Equivalence
Legend: P Stride Equivalence
Validation Publications:
Duan B, He L, Zhang J, Fang L, Li G. Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020). Blood Press Monit. 2023 Oct 1;28(5):284-288. Epub: 2023 Jul 12. doi: 10.1097/MBP.0000000000000662. PMID: 37434533. Available from: PMC10484188.
81060-2:2018/Amd 1:2020 - Pass General population
Yang R, Li F. Validation of the DBP-6279B arm-type fully automatic digital blood pressure monitor in pregnancy according to the ISO 81060-2:2018/AMD 1:2020 protocol. Blood Press Monit. Epub: 2025 Nov 6. doi: 10.1097/MBP.0000000000000779. PMID: 41190939.
81060-2:2018/Amd 1:2020 - Pass Pregnancy