Philips IntelliVue MP50 (M8004A)
Device Name:
IntelliVue MP-50
Device Model:
M8004A
Manufacturer:
Philips Electronics North America Corp., 3000 Minuteman Road, M/S 109, Andover, MA 01810, UNITED STATES.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for patient monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Continuous measurement
Availability:
Available Currently
Device Manual:
Description:
The Philips IntelliVue MP50 (M8004A) is a patient monitor. Medaval has not found evidence proving the accuracy of its blood pressure measurement technology. Blood pressure measurements are taken from the upper arm. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Philips IntelliVue MP50 (M8004A), to measure blood pressure, has failed in a clinical validation study, in a specific population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
| Accuracy Assessment | Recommendation | Basis |
BP |
Medaval |
Not recommended |
This device has failed a clinical validation. |
Validation Publications:
Meng X, Zang G, Fan L, Zheng L, Dai J, Wang X, Xia W, Liu J, Zhang C. Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia. Exp Ther Med. 2013 Jul;6(1):9-14. Epub: 2013 May 17. doi: 10.3892/etm.2013.1121. PMID: 23935710. Available from: PMC3735910.
IA protocol - Fail (n=515)
Meng X, Zang G, Li R, Fan L, Liu J, Wang Z. Study of the disparity of noninvasive and invasive blood pressure measured by the Philips Intellivue MP50 monitor in surgeries inducing gall cardiac reflex under general anesthesia. Blood Press Monit. 2018 Feb;23(1):19-23. Epub: 2017 Nov 08. doi: 10.1097/MBP.0000000000000304. PMID: 29120887.
IA protocol - Fail (n=78)