Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
BG | MDR Criteria | Self-measurement and professional use | Published evidence |
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.
15197:2003 - Fail General population
15197:2013 - Fail General population
Brazg RL, Klaff LJ, Parkin CG. Performance variability of seven commonly used self-monitoring of blood glucose systems: clinical considerations for patients and providers. J Diabetes Sci Technol. 2013 Jan 1;7(1):144-52. doi: 10.1177/193229681300700117. PMID: 23439170. Available from: PMC3692226.
15197:2003 - Pass General population
Jeanny G, Hope P. Surveillance of the System Accuracy of Two Systems for Self-Monitoring of Blood Glucose After Market Approval. J Diabetes Sci Technol. 2016 Jan;10(1):240-1. Epub: 2015 Sep 30. doi: 10.1177/1932296815608871. PMID: 26424240. Available from: PMC4738221. No abstract available.
15197:2013 - Pass General population
15197:2003 - Pass General population
Park HI, Lee SS, Son JW, Kwon HS, Kim SR, Chae H, Kim M, Kim Y, Yoo S. Analytical Performance Evaluation of Infopia Elementâ„¢ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose. J Clin Lab Anal. 2016 Nov;30(6):449-58. Epub: 2016 Apr 13. doi: 10.1002/jcla.21947. PMID: 27074748.
15197:2013 - Pass General population
Author's Note: Elementâ„¢ BGMS was considered an appropriate SMBG for home use; however, the positive bias at low-to-mid glucose levels requires further improvement.