Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
Legend: B BIHS Derivative, D DET Equivalence
Ware A, Stevens R, Selwood M, Fleming S. Accuracy of blood pressure monitors available in high street pharmacies. Blood Press Monit. 2016 Feb;21(1):59-61. doi: 10.1097/MBP.0000000000000151. PMID: 26427055.
Measurements from the Omron R7 (HEM-637-E7) are compared to those from the Omron PA-350 (BPM tester), though not in a formal validation.
Halfon M, Wuerzner G, Marques-Vidal P, Taffe P, Vaucher J, Waeber B, Liaudet L, Ltaief Z, Popov M, Waeber G. Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement. Blood Press. 2018 Feb;27(1):48-55. Epub: 2017 Sep 28. doi: 10.1080/08037051.2017.1383852. PMID: 28958152.
Measurements from the Omron R7 (HEM-637-E7) are compared to those from the Microlife WatchBP Home (BP 3MX1-1) and the Omron HEM-907 (HEM-907-E7) in atrial fibrillation, though not in a formal validation.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron 637IT (HEM-637IT-E)
Altunkan S, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly. Blood Press Monit. 2006 Apr;11(2):97-102. PMID: 16534412.
ESH-IP:2002 - Pass Elderly (Age>65)
Altunkan S, OztaĆ K, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults. Blood Press Monit. 2006 Apr;11(2):79-85. PMID: 16534409.
ESH-IP:2002 - Pass General population
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit. 2006 Jun;11(3):165-71. PMID: 16702826.
ESH-IP:2002 - Pass General population
Omron R2 (HEM-6113-E)
Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: PMC3237100.
ESH-IP:2002 - Pass General population