Omron MIT (HEM-751-E)
Device Name:
MIT
Device Model:
HEM-751-E
Manufacturer:
Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Availability according to Countries or Regions:
Germany
Description:
The Omron MIT (HEM-751-E) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The technology used in the Omron MIT (HEM-751-E), to measure blood pressure, has passed in six clinical validation studies, between general and pregnancy populations, according to recognised standard protocols, as published in peer-reviewed publications.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommendation Older clinical validation; older protocol; multiple clinical validations
BP BIHS (UK and IRL) Previous recommendation, now archived for pregnancy (Archived) Published evidence
BP DHL (DE) General use (Previous: 2001) Published evidence
BP ESH (Europe) Self-measurement and professional use for adults and pregnancy Published evidence
BP Stride BP Self-measurement for adults and pregnancy (Unavailable) Published evidence
Validation Publications:

Golara M, Jones C, Randhawa M, Shennan AH. Inflationary oscillometric blood pressure monitoring: validation of the OMRON-MIT. Blood Press Monit. 2002 Dec;7(6):325-8. PMID: 12488653.

BHS:1993 - Pass (A/A) General population

SP10:1992 - Pass General population

Golara M, Benedict A, Jones C, Randhawa M, Poston L, Shennan AH. Inflationary oscillometry provides accurate measurement of blood pressure in pre-eclampsia. BJOG. 2002 Oct;109(10):1143-7. doi: 10.1111/j.1471-0528.2002.01487.x. PMID: 12387468. Available from: obgyn.onlinelibrary.wiley.com.

BHS:1993 - Pass (B/A) Pregnancy

SP10:1987 - Pass Pregnancy

BHS:1993 - Pass (B/A) Pre-eclampsia

SP10:1987 - Pass Pre-eclampsia

de Greeff A, Beg Z, Gangji Z, Dorney E, Shennan AH. Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7. Blood Press Monit. 2009 Feb;14(1):37-40. doi: 10.1097/MBP.0b013e32831e305d. PMID: 19252437.

BHS:1993 - Pass (A/A) Pregnancy (n=30)

SP10:1992 - Pass Pregnancy (n=30)

BHS:1993 - Pass (A/A) Pre-eclampsia (n=15)

SP10:1992 - Pass Pre-eclampsia (n=15)

Tholl U, Lüders S, Bramlage P, Dechend R, Eckert S, Mengden T, Nürnberger J, Sanner B, Anlauf M. The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control. Blood Press Monit. 2016 Aug;22(4):197-205. doi: 10.1097/MBP.0000000000000186. PMID: 26998590.

DHL:1999 - Pass General population (Note: 2001)