Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★★ Recommendation | Recent clinical validation; recent protocol |
BP | Stride BP | Self-measurement for adults (Unavailable) | Published evidence |
BP | MDR Criteria | Self-measurement | Published evidence |
Kuwabara M, Harada K, Hishiki Y, Kario K. Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T. J Clin Hypertens (Greenwich). 2019 Apr;21(4):463-469. Epub: 2019 Jan 4. doi: 10.1111/jch.13464. PMID: 30609129. Available from: onlinelibrary.wiley.com.
81060-2:2013 - Pass Sitting position
81060-2:2013 - Fail Supine position, palm sideways
81060-2:2013 - Fail Supine position, palm upwards
81060-2:2013 - Pass Supine position, palm downwards
This study indicates that the Omron HEM-9600T is suitable for blood pressure measurement in the seated position, with the wrist positioned correctly. However, it is not suitable for its intended use in measuring nocturnal pressure.