Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | Recommended in certain condition(s) | Validation in specific condition(s). See publications for details. |
BP | ESH (Europe) | Self-measurement for adults and children | Published evidence |
BP | Japanese Society of Hypertension | Self-measurement (Previous: 2016-2018) | Implied equivalence to the Omron 705IT (HEM-759-E), M5-I (HEM-757-E) or T9P |
BP | Stride BP | Self-measurement for adults and children (Unavailable) | Equivalence claim without published evidence |
Legend: B BIHS Derivative, D DET Equivalence, J JHS Implied Equivalence, P Stride Equivalence
Godai K, Kabayama M, Saito K, Asayama K, Yamamoto K, Sugimoto K, Ohkubo T, Rakugi H, Kamide K. Validation of an automated home blood pressure measurement device in oldest-old populations. Hypertens Res. 2020 Jan;43(1):30-35. Epub: 2019 Sep 18. doi: 10.1038/s41440-019-0330-7. PMID: 31534190.
81060-2:2013 - Pass Very elderly, aged 85+ years
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron 705IT (HEM-759-E)
El Assaad MA, Topouchian JA, Asmar RG. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit. 2003 Jun;8(3):127-33. PMID: 12900590.
ESH-IP:2002 - Pass General population
Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: the Arsakion School Study. Blood Press Monit. 2006 Aug;11(4):229-34. PMID: 16810034.
ESH-IP:2002 - Pass Children and adolescents (n=197)
SP10:1992 - Pass Children and adolescents (n=197)
Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit. 2006 Feb;11(1):27-32. PMID: 16410738.
BHS:1993 - Pass (A/A) General population
SP10:2002 - Pass General population
Omron M6 (HEM-7001-E)
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit. 2006 Jun;11(3):165-71. PMID: 16702826.
ESH-IP:2002 - Pass General population
Altunkan S, Ilman N, Kayatürk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults. Blood Press Monit. 2007 Aug;12(4):219-25. PMID: 17625394.
ESH-IP:2002 - Pass Adults
ESH-IP:2002 - Pass Obese adults
Altunkan S, Iliman N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients. Blood Press Monit. 2008 Apr;13(2):117-22. doi: 10.1097/MBP.0b013e3282f3fb50. PMID: 18347447.
ESH-IP:2002 - Pass Elderly
Omron M6 (HEM-7211-E)
Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: PMC3237100.
ESH-IP:2010 - Pass General population