Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations |
BP | ESH (Europe) | Self-measurement | Published evidence |
BP | Stride BP | Self-measurement for adults (Unavailable) | Published evidence |
BP | FDA (US) | 510(k) | 510(k) K914823 |
Foster C, McKinlay S, Cruickshank JM, Coats AJ. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens. 1994 Sep;8(9):661-4. PMID: 7807495.
BHS:1993 - Fail (B/C) General population
SP10:1992 - Pass General population
Artigao LM, Llavador JJ, Puras A, López Abril J, Rubio MM, Torres C, Vidal A, Sanchis C, Divisón JA, Naharro F, Caldevilla D, Fuentes G. Evaluación y validación de los monitores Omron Hem 705 CP y Hem 706/711 para automedidas de presión arterial [Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure.] Aten Primaria. 2000 Feb 15;25(2):74-84. PMID: 10736939. Available from: www.elsevier.es. In Spanish.
BHS:1993 - Pass (B/A) General population
SP10:1996 - Pass General population