Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | Japanese Society of Hypertension | Self-measurement (Previous: 2023) | Manufacturer declaration of equivalence to the Omron HEM-1020 |
Legend: J JHS Implied Equivalence
Liang Z, Chapa-Martell MA. Validation of Omron Wearable Blood Pressure Monitor HeartGuideā¢ in Free-Living Environments. In: Wireless Mobile Communication and Healthcare: 9th International Conference, MobiHealth 2020, Virtual Event, November 19, 2020, Proceedings. Eds: Ye J, O'Grady MJ, Civitarese G, Yordanova K. Basel, Switzerland: Springer Nature; 2021 Mar 24. p. 339-350.
Measurements from the Omron HEM-1022 are compared to those from the Omron Heartguide BP8000-M (HEM-6410T-ZM) or the BP8000-L (HEM-6410T-ZL), though not in a formal validation.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron HEM-1020
Takahashi H, Yoshika M, Yokoi T. Validation of home blood pressure-monitoring devices, Omron HEM-1020 and Omron i-Q132 (HEM-1010-E), according to the European Society of Hypertension International Protocol. Blood Press Monit. 2011 Aug;16(4):203-7. doi: 10.1097/MBP.0b013e328348b688. PMID: 21701384.
ESH-IP:2002 - Pass General population