📞  +353 1 6643768
 
📧  info@medaval.ie
Omron BP5450 (HEM-7343T-Z) Platinum
Device Name:
BP5450 Platinum
Device Model:
HEM-7343T-Z
Manufacturer:
Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
Canada and United States
Description:
The Omron BP5450 (HEM-7343T-Z) Platinum is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Omron BP5450 (HEM-7343T-Z) Platinum, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★ Recommendation Recent clinical validation; recent protocol
BP Hypertension Canada Self-measurement (Gold) Manufacturer declaration of equivalence
BP Stride BP Self-measurement for adults (Preferred) Published evidence
BP AMA VDL (US) Self-measurement Implied equivalence to the Omron HEM-7311
Consumer BodyYear(s)LevelComment
BP Consumer Reports (US) Latest: 2021 Excellent
Validation Publications:

Peprah YA, Lee JY, Persell SD. Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol. J Hum Hypertens. 2023 Feb;37(2):134-140. Epub: 2023 Jan 18. doi: 10.1038/s41371-022-00795-6. PMID: 36653402. Available from: PMC9957726.

81060-2:2020 - Fail (n=85) (Note: Marginal DBP fail for Criterion 2)

81060-2:2020 - Pass (n=86) (Note: Marginal DBP pass for Criterion 2 after extra subject added)