Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations |
BP | BIHS (UK and IRL) | Previous recommendation, now archived for adults, obese adults and the elderly (Archived) | Published evidence |
BP | ESH (Europe) | Self-measurement | Published evidence |
BP | Stride BP | Self-measurement for adults, obese adults and the elderly (Unavailable) | Published evidence, but the device validated was the |
Legend: B BIHS Derivative, D DET Equivalence
Altunkan S, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly. Blood Press Monit. 2006 Apr;11(2):97-102. PMID: 16534412.
ESH-IP:2002 - Pass Elderly (Age>65)
Altunkan S, Oztaş K, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults. Blood Press Monit. 2006 Apr;11(2):79-85. PMID: 16534409.
ESH-IP:2002 - Pass General population
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit. 2006 Jun;11(3):165-71. PMID: 16702826.
ESH-IP:2002 - Pass General population
Altunkan S, Genç Y, Altunkan E. A comparative study of an ambulatory blood pressure measuring device and a wrist blood pressure monitor with a position sensor versus a mercury sphygmomanometer. Eur J Intern Med. 2007 Mar;18(2):118-23. PMID: 17338963.
Measurements from the Omron 637IT (HEM-637IT-E) are compared to those from the Nissei DS-250 and a standard mercury sphygmomanometer, though not in a formal validation.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron R2 (HEM-6113-E)
Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: PMC3237100.
ESH-IP:2002 - Pass General population