Nihon-Kohden PVM-2701

Device Name:
Vismo PVM-2700 Series
Device Model:
PVM-2701
Manufacturer:
Distributer: Nihon Kohden UK Ltd., Unit 118 Trident Court, 1 Oakcroft Road Chessington, Surrey KT9 1BD, UNITED KINGDOM.
Other: Nihon Kohden Corporation, 1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, JAPAN.
Measuring functions:
Blood Pressure, SpO2, PR, ECG, Temperature, Respiration and Arrhythmias
Primary Client Use:
Intended for patient monitoring
Measurement Site:
BP: Upper Arm; SpO2: Finger
Measurement Occurrence:
BP: Single and intermittent measurements; SpO2: Continuous measurement
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Device Specifications:
Description:
The Nihon-Kohden PVM-2701 is a vital signs monitor. Its blood pressure measurement technology has been proven to be accurate and information regarding the accuracy of its oxygen saturation measurement technology does not appear to be available. Blood pressure measurements are taken from the upper arm and oxygen saturation measurements are taken from the finger. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Nihon-Kohden PVM-2701, to measure blood pressure, has passed in a clinical validation study, in a general population including children, according to a recognised standard protocol, as published in a peer-reviewed publication. There appears to be no peer-reviewed clinical validation information available on the technology used in the Nihon-Kohden PVM-2701 to measure peripheral oxygen saturation.
Recommendations:
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | Legacy approval | Older clinical validation; older protocol |
BP | ESH (Europe) | Professional use for children | Published evidence |
BP | Japanese Society of Hypertension | Professional use (2019-2024) | Implied equivalence to the Nihon Kohden PVM-2701/Impulse-1 |
BP | Stride BP | Hospital use for adults and children | Published evidence |
Device Family:
Nihon-Kohden PVM-2701J, Nihon-Kohden PVM-2703J, Nihon-Kohden BSM-1700 (PT)J, Nihon-Kohden BSM-3500 Series (VS)J, Nihon-Kohden BSM-6000 Series (TR)J, Nihon-Kohden CSM-1500 Series (G5)J, Nihon-Kohden CSM-1700 Series (G7)J, Nihon-Kohden CSM-1900 Series (G9)J, Nihon-Kohden BSM-3700 Series (VS)J, Nihon-Kohden TEC-8300 SeriesJ, Nihon-Kohden ZM-540J, Nihon-Kohden ZM-940PGJ, Nihon-Kohden GZ-140PJ, Nihon-Kohden PVM-9900J, Nihon-Kohden RMC-5000J, Nihon-Kohden SVM-7000 SeriesJ, Nihon-Kohden TEC-5600 SeriesJ, Nihon-Kohden ZS-640J, Nihon-Kohden ZS-940PJ, Nihon-Kohden PVM-4000J, Nihon-Kohden EMS-1052J, Nihon-Kohden BSM-5700 Series (E7)J, Nihon-Kohden TEC-1000 SeriesJ
Legend: J JHS Implied Equivalence
Legend: J JHS Implied Equivalence
Validation Publications:
Alpert BS. Validation of the Nihon Kohden PVM-2701/Impulse-1 automated device by both AAMI (2002) and ISO standards testing. Blood Press Monit. 2012 Oct;17(5):207-9. doi: 10.1097/MBP.0b013e328359c48e. PMID: 22968109.
81060-2:2009 - Pass Adults (n=69) and children (n=41)
SP10:2002 - Pass Adults (n=69) and children (n=41)