Nihon-Kohden Vismo PVM-2701
Device Name:
Vismo
Device Model:
PVM-2701
Manufacturer:

Distributer: Nihon Kohden UK Ltd., Unit 118 Trident Court, 1 Oakcroft Road Chessington, Surrey KT9 1BD, UNITED KINGDOM.

Other: Nihon Kohden Corporation, 1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, JAPAN.

Measuring functions:
Blood pressure, Arrhythmias, SpO2, Temperature, ECG, Respiration
Primary Client Use:
Intended for patient monitoring
Measurement Site:
BP: Upper Arm; SpO2: Finger
Measurement Occurrence:
BP & SpO2: Continuous measurement
Availability:
Available Currently
Device Specifications:
Description:
The Nihon-Kohden Vismo PVM-2701 is a patient monitor. Its blood pressure measurement technology has been proven to be accurate and information regarding the accuracy of its oxygen saturation measurement technology does not appear to be available. Blood pressure measurements are taken from the upper arm and oxygen saturation measurements are taken from the finger. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Nihon-Kohden Vismo PVM-2701, to measure blood pressure, has passed in a clinical validation study, in a general population including children, according to a recognised standard protocol, as published in a peer-reviewed publication. There appears to be no peer-reviewed clinical validation information available on the technology used in the Nihon-Kohden Vismo PVM-2701 to measure peripheral oxygen saturation.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Legacy approval Older clinical validation; older protocol
BP ESH (Europe) Professional use for children Published evidence
BP Stride BP Hospital use for adults and children Published evidence
Device Family:
Validation Publications:

Alpert BS. Validation of the Nihon Kohden PVM-2701/Impulse-1 automated device by both AAMI (2002) and ISO standards testing. Blood Press Monit. 2012 Oct;17(5):207-9. doi: 10.1097/MBP.0b013e328359c48e. PMID: 22968109.

81060-2:2009 - Pass Adults (n=69) and children (n=41)

SP10:2002 - Pass Adults (n=69) and children (n=41)