Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | ★ Recommendation | Older clinical validation; older protocol; multiple clinical validations |
BP | BIHS (UK and IRL) | Professional use for pregnancy | Published evidence |
BP | ESH (Europe) | Self-measurement and professional use for adults and pregnancy | Published evidence |
BP | Stride BP | Office use for adults and pregnancy (Unavailable) (Preferred) | Published evidence |
BP | MDR Criteria | Pregnancy | Published evidence |
de Greeff A, Nathan H, Stafford N, Liu B, Shennan AH. Development of an accurate oscillometric blood pressure device for low resource settings. Blood Press Monit. 2008 Dec;13(6):342-8. doi: 10.1097/MBP.0b013e32830fd07c. PMID: 19020425.
ESH-IP:2002 - Pass General population
Nathan HL, de Greeff A, Hezelgrave NL, Chappell LC, Shennan AH. An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2. Blood Press Monit. 2015 Feb;20(1):52-5. doi: 10.1097/MBP.0000000000000086. PMID: 25243711.
BHS:1993 - Pass (B/A) Pregnancy and pre-eclampsia (n=45)
81060-2:2013 - Pass Pregnancy and pre-eclampsia (n=45)
Nathan HL, de Greeff A, Hezelgrave NL, Chappell LC, Shennan AH. Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure. Blood Press Monit. 2015 Oct;20(5):299-302. Epub: 2015 May 5. doi: 10.1097/MBP.0000000000000134. PMID: 26020367.
BHS:1993 - Pass (A/A) Low BP pregnancy (n=30)
81060-2:2013 - Pass Low BP pregnancy (n=30)