Medipro MP-BP-BPM
Device Name:
MP-BP-BPM
Manufacturer:

OBL: Shanghai Medipro Medical Co. Ltd., 6018 Huyi Road, Waigang, Jiading District, Shanghai, CHINA.

OEM: Shanghai Honsun International Trade Co. Ltd., Room C. Floor 22, Zhiyuan Building, 768 Xietu Road, Shanghai 200023, CHINA.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The Medipro MP-BP-BPM is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for professional use.
Assessment:
While the technology used in the Medipro MP-BP-BPM, to measure blood pressure, has been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Self-measurement Manufacturer declaration of equivalence to the Honsun LD-533
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Honsun Scian LD-578

Zhang Y, Wang J, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit. 2009 Jun;14(3):128-31. doi: 10.1097/MBP.0b013e32832ce149. PMID: 19454866.

ESH-IP:2002 - Pass General population