Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★ Recommendation | Older clinical validation; recent protocol; multiple clinical validations |
BG | MDR Criteria | Self-measurement and professional use | Published evidence |
Consumer Body | Year(s) | Level | Comment | |
BG | Amazon Best Sellers (US) | Latest: 2019 | A (3.9 / 5) |
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.
15197:2003 - Pass General population
15197:2013 - Pass General population
Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: PMC3876330.
15197:2003 - Pass Using insufficient blood samples (n=31)
Sølvik UØ, Risa M, Jacobsen CE, Monsen G, Sandberg S. Performance of 10 Systems for Self-Monitoring of Blood Glucose by Trained Healthcare Professionals and in the Hands of the Users. Clin Chem. 2015 May;61(5):772-4. Epub: 2015 Mar 3. doi: 10.1373/clinchem.2014.236760. PMID: 25737533. Available from: clinchem.aaccjnls.org.
15197:2013 - Pass General population (Note: BLS-recorded measurements)
FDA:2016H - Pass General population (Note: BLS-recorded measurements)
15197:2013 - Fail General population (Note: Subject-recorded measurements)
FDA:2016H - Fail General population (Note: Subject-recorded measurements)
15197:2003 - Pass General population
SKUP Evaluation 2011/86
Jendrike N, Baumstark A, Pleus S, Liebing C, Kamecke U, Haug C, Freckmann G. Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. Curr Med Res Opin. 2019 Feb;35(2):301-311. Epub: 2018 Jul 19. doi: 10.1080/03007995.2018.1491832. PMID: 29927663.
15197:2013 General population (Note: Results not in Abstract)
Klonoff DC, Parkes JL, Kovatchev BP, Kerr D, Bevier WC, Brazg RL, Christiansen M, Bailey TS, Nichols JH, Kohn MA. Investigation of the Accuracy of 18 Marketed Blood Glucose Monitors. Diabetes Care. 2018 Aug;41(8):1681-8. Epub: 2018 Jun 13. doi: 10.2337/dc17-1960. PMID: 29898901. Available from: www.diabetestechnology.org.
DTS:2016 - Fail General population
SKUP. Report from the evaluation SKUP/2011/86: OneTouch Verio (LifeScan, Johnson & Johnson), a system for blood glucose measurement. Bergen, Norway: SKUP; 2011. 67 p. Available from: skup.org.
15197:2003 - Pass General population
Results counted in Sølvik et al. 2015.
Musholt PB, Schipper C, Thomé N, Ramljak S, Schmidt M, Forst T, Pfützner A. Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices. J Diabetes Sci Technol. 2011 Sep 1;5(5):1167-75. doi: 10.1177/193229681100500520. PMID: 22027312. Available from: PMC3208875.
Reliable correction for potential haematocrit influence.
Littmann K, Petersen ER, Pussinen C, Danielson K, Djurisic S, Eilertsen H, Garabet L, Greibe E, Lauritzen T, Olsen DA, Othman S, Palimaru I, Westerlund J. Evaluation of OneTouch Verio®, a new blood glucose self-monitoring system for patients with diabetes. Scand J Clin Lab Invest. 2013;73(4):286-92. Epub: 2013 Mar 11. doi: 10.3109/00365513.2013.773064. PMID: 23473019.
This is a supplementary presentation of validation results for the LifeScan OneTouch Verio (V1).
SKUP Evaluation 2011/86. See: Sølvik et al. 2015
Teodorczyk M, Nandagopalan S, Maguire P, Stegmann J. System accuracy of blood glucose monitoring devices according to the current and proposed ISO 15197 standards. J Diabetes Sci Technol. 2013 May 1;7(3):795-7. doi: 10.1177/193229681300700325. PMID: 23759413. Available from: PMC3869148.
The publication is a letter or correspondence in which the LifeScan OneTouch Verio (V1) is referenced.
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. In response to Teodorczyk and coauthors: System accuracy of blood glucose monitoring devices according to the current and proposed ISO 15197 standards. J Diabetes Sci Technol. 2013 Nov 1;7(6):1659-60. doi: 10.1177/193229681300700632. PMID: 24351198. Available from: PMC3876350.
The publication is a letter or correspondence in which the LifeScan OneTouch Verio (V1) is referenced.
Hwang JH, Sohn YH, Chang SS, Kim SY. Evaluation of three glucometers for whole blood glucose measurements at the point of care in preterm or low-birth-weight infants. Korean J Pediatr. 2015 Aug;58(8):301-308. Epub: 2015 Aug 21. doi: 10.3345/kjp.2015.58.8.301. PMID: 26388895. Available from: PMC4573444.
The publication evaluates the performance of the LifeScan OneTouch Verio (V1), though not in a formal validation.