Heine Gamma XXL LF (Stand)
Device Name:
Gamma XXL LF
Device Model:
M-000.09.326
Manufacturer:
HEINE Optotechnik GmbH & Co. KG, Kientalstraße 7, 82211 Herrsching a. Ammersee, GERMANY.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Device Specifications:
Description:
The Heine Gamma XXL LF (Stand) is a mobile (floor, stand, trolly) manually-operated aneroid sphygmomanometer. Blood pressure measurements are taken from the upper arm. It is intended for professional use.
Assessment:
There appears to be no peer-reviewed clinical validation information available on the technology used in the Heine Gamma XXL LF (Stand) to measure blood pressure. However, this device appears to be very similar to a validated device by the same manufacturer or a device within a family of equivalent devices, but equivalence needs to be proven according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None This is a manual device. Equivalence to a validated device is possible but this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Professional use Published evidence
BP ESH (Europe) Professional use Published evidence
Device Family:
Heine Gamma XXL LF (Desk)Z, Heine Gamma XXL LF (Wall)Z, Heine Gamma XXL LF (Rail)Z, Heine Gamma XXL LF (Stand)Z
Legend: Z Almost Certain
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Heine Gamma XXL LF (Desk)

Dorigatti F, Bonso E, Zanier A, Palatini P. Validation of Heine Gamma G7 (G5) and XXL-LF aneroid devices for blood pressure measurement. Blood Press Monit. 2007 Feb;12(1):29-33. PMID: 17303985.

ESH-IP:2002 - Pass General population