📞  +353 1 6643768
 
📧  info@medaval.ie
H&L HL868ED
Device Model:
HL868ED
Manufacturer:
Health & Life Co. Ltd., 9F - 186 Jian 1st Rd., Zhonghe Dist., New Taipei City 23553, TAIWAN ROC.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The H&L HL868ED is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the H&L HL868ED, to measure blood pressure, has passed in two clinical validation studies, in general populations, according to recognised standard protocols, as published in peer-reviewed publications.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★ Recommendation Recent clinical validation; older protocol; multiple clinical validations
BP BIHS (UK and IRL) Self-measurement Published evidence
BP Stride BP Self-measurement for adults (Preferred) Published evidence
Device Family:
H&L HL858GT TalkingP, H&L HL868EDP, H&L HL858A2-TP, Seagull S-868EDP
Legend: P Stride Equivalence
Validation Publications:

Zhang W, Lei L, Li Y, Wang JG. Validation of the HL868ED upper-arm blood pressure monitor for clinical use and self-measurement according to the British Hypertension Society Protocol. Blood Press Monit. 2019 Aug;24(4):203-207. Epub: 2019 May 23. doi: 10.1097/MBP.0000000000000384. PMID: 31136322. Erratum in: Blood Press Monit. 2021 Feb;26(1):85.

BHS:1993 - Pass (A/A) General population

Zhang W, Lei L, Li Y, Wang JG. Validation of the HL868ED upper-arm blood pressure monitor for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2020 Feb;25(1):53-57. Epub: 2019 Oct. doi: 10.1097/MBP.0000000000000411. PMID: 31658108. Erratum in: Blood Press Monit. 2021 Feb;26(1):84.

ESH-IP:2010 - Pass General population