H&L HL868BA
Device Model:
HL868BA
Manufacturer:
Health & Life Co. Ltd., 9F - 186 Jian 1st Rd., Zhonghe Dist., New Taipei City 23553, TAIWAN ROC.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Device Manual:
Description:
The H&L HL868BA is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the H&L HL868BA, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Legacy approval Older clinical validation; older protocol
BP BIHS (UK and IRL) Self-measurement (Previous) Published evidence
BP Stride BP Self-measurement for adults (Unavailable) Published evidence
Device Family:
Validation Publications:

Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit. 2008 Oct;13(5):305-8. doi: 10.1097/MBP.0b013e32830d4bd0. PMID: 18799958.

ESH-IP:2002 - Pass General population

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Sejoy Joytech DBP-1307 (Sejoy BP-1307, Joytech DBP-1307)

Lei L, Chen Y, Chen Q, Li Y, Wang JG. Validation of the SEJOY BP-1307 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2017 Dec;22(6):371-4. Epub: 2017 Aug 03. doi: 10.1097/MBP.0000000000000283. PMID: 28786853 .

ESH-IP:2010 - Pass General population