Gambro Dasco HD-BPM
Device Name:
HD-BPM
Manufacturer:
Gambro Dasco S.p.A. (Baxter), Via Modenese 66, IT-43136 Medolla (MO), ITALY.
Measuring functions:
Blood pressure
Primary Client Use:
Module intended as a component of a patient monitor
Measurement Site:
Upper Arm
Measurement Occurrence:
Continuous measurement
Availability:
Obsolete
Description:
The Gambro Dasco HD-BPM is a blood pressure module for a patient monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for use in a bedside patient monitor. This device is now obsolete.
Assessment:
The technology used in the Gambro Dasco HD-BPM, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | Legacy approval | Older clinical validation; older protocol |
BP | Stride BP | Hospital use for adults | Published evidence |
Validation Publications:
Cavalcanti S, Marchesi G, Ghidini C. Validation of automated oscillometric sphygmomanometer (HDBPM) for arterial pressure measurement during haemodialysis. Med Biol Eng Comput. 2000 Jan;38(1):98-101. PMID: 10829398.
SP10:1992 - Pass General population
Relevant Publications:
Lodi CA, Estridge C, Ghidini C. In vitro and in vivo evaluation of an oscillometric device for monitoring blood pressure in dialysis patients. Nephrol Dial Transplant. 2007 Oct;22(10):2950-61. Epub: 2007 Jun 7. doi: 10.1093/ndt/gfm288. PMID: 17556423 .
This study was carried out by the manufacturers. The electronic measurement-data from the 2000 study was used to check a revised version of the device. It passed. Supplementary comparisons with a simulator and the Colin BX-100 were also performed.