Device Name:
Fora MD6
Device Model:
TD-4172B
Manufacturer:

OBL: Puyang Technology Co. Ltd. (ForaCare), 1st Floor, 127 Wugong 2nd Road, Wugu District 24888, New Taipei City, TAIWAN ROC.

OEM: TaiDoc Technology Corporation, 6F - 127 Wugong 2nd Road, 24888 Wugu District, New Taipei City, TAIWAN ROC.

Measuring functions:
Blood glucose
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Finger and general Alternative Site Testing
Measurement Occurrence:
Single measurements only
Connected Health Technologies:
Bluetooth
Availability:
Available Currently
Availability according to Countries or Regions:
Taiwan ROC
Description:
The Fora MD6 (TD-4172B) is an automatic blood glucose meter. Medaval has not found evidence proving the accuracy of its blood glucose measurement technology. Blood glucose measurements are normally taken from the finger but alternative site testing is allowed. It is intended for self-measurement and home use.
Assessment:
There appears to be no peer-reviewed clinical validation information available on the technology used in the Fora MD6 (TD-4172B) to measure blood glucose. However, this device appears to be very similar to another device by the same manufacturer or a device within a family of equivalent devices. At least one validation needs to be published and equivalence needs to be proven according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BG Medaval Not recommended This device has not been shown to have been clinically validated.